FLOGARD
Report
- Report Number
- 1416980-2014-16963
- Event Type
- Malfunction
- Date Received
- May 25, 2014
- Date of Event
- April 30, 2014
- Report Date
- April 30, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). (B)(6). EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. VISUAL INSPECTION, FUNCTIONAL TESTING, BATTERY TESTING, AND A REVIEW OF THE ALARM LOG WERE PERFORMED. FUNCTIONAL TESTING AND THE REVIEW OF THE ALARM LOG REVEALED THAT THE DEVICE HAD PRESENTED AN F-38 ALARM, CONFIRMING THE REPORTED CONDITION. THE CAUSE OF THE F-38 ALARM WAS DETERMINED TO BE OUT OF SPECIFICATION FORCE SENSING RESISTORS (FSRS). TO CORRECT THE CONDITION, THE FSRS WERE REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A FLOGARD INFUSION PUMP EXPERIENCED THE F-38 ALARM. THERE WAS NO PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 309761 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |