HEARTWARE® VENTRICULAR ASSIST SYSTEM
Report
- Report Number
- 3007042319-2014-00539
- Event Type
- Malfunction
- Date Received
- May 25, 2014
- Date of Event
- September 3, 2013
- Report Date
- April 25, 2014
- Manufacturer
- HEARTWARE, INC
- Product Code
- DSQ
- PMA / PMN Number
- P100047
- Removal / Correction Number
- Z-1607-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
PRELIMINARY EVALUATION AND TESTING REVEALED THAT THE RETURNED BATTERY CONTAINED A FAULTY CELL. THIS FINDING WAS RE-ASSESSED PER OUR SOP REQUIREMENTS, AND DETERMINED TO BE REPORTABLE. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN THIRTY (30) DAYS OF COMPLETION OF THE INVESTIGATION. EVALUATION IN PROGRESS.
THE BATTERY WAS RETURNED; VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICE IN RELATION TO THE REPORTED EVENT. THOROUGH EXTERNAL VISUAL INSPECTION OF THE BATTERY REVEALED NO SIGNS OF PHYSICAL DAMAGE OR CONTAMINATION. FUNCTIONAL TESTING REVEALED THE RETURNED BATTERY CONTAINED A FAULTY CELL PAIR, WHICH MOST LIKELY CONTRIBUTED TO THE REPORTED EVENT. AN INTERNAL INVESTIGATION (CAPA) HAS BEEN OPENED TO ADDRESS THE ISSUE. THIS IS ONE OF FOUR REPORTS (3007042319-2014-00539, 00594, 00595, AND 2014-00596) SUBMITTED FOR DEVICES RELATED TO THE SAME EVENT.
APPROXIMATELY ONE YEAR POST HVAD IMPLANTATION THE SITE REPORTED THAT THE PATIENT EXPERIENCED A CHANGE OF POWER SOURCE IN THE CONTROLLER EARLIER THAN EXPECTED. THE BATTERY WAS REMOVED FROM THE PATIENT AND A NEW BATTERY WAS SUPPLIED. NO HARM OR INJURY TO THE PATIENT WAS REPORTED AS A RESULT OF THIS INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 309760 | HEARTWARE® VENTRICULAR ASSIST SYSTEM | CIRCULATORY ASSIST SYSTEM, BATTERY | DSQ | HEARTWARE, INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | (B)(4) - BATTERY| (B)(4) - BATTERY| (B)(4) - BATTERY| (B)(4) - BATTERY |