FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3831367 · Received May 24, 2014

Report

Report Number
2531779-2014-14734
Event Type
Malfunction
Date Received
May 24, 2014
Report Date
May 19, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/19/2014 WITH THE FOLLOWING FINDINGS:INVESTIGATION REVEALED THAT THE DISPLAY SCREEN WAS DIM AND DISCOLORED. UNRELATED TO THE DISPLAY ISSUE, EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PRINTED CIRCUIT BOARD HAD FAILED. THE PUMP WOULD NOT RETAIN THE USER PROGRAMMED DATE AND TIME SETTINGS UPON REMOVAL OF THE PRIMARY AA BATTERY. WHEN A NEW AA BATTERY IS INSERTED THE PUMP DISPLAYS THE DEFAULT DATE AND TIME WHICH MUST BE MANUALLY CONFIRMED (OR RESET) BY THE USER IN ORDER TO PROCEED. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DISPLAY ISSUE. THIS REPORT IS MADE BASED ON RESULTS OF THE INVESTIGATION PERFORMED ON 05/19/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
309606 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 62 YR