ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-14722
- Event Type
- Malfunction
- Date Received
- May 24, 2014
- Report Date
- May 20, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/17/2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX SHOWED UNEXPLAINED REBOOTS. A VISUAL INSPECTION SHOWED THE BATTERY COMPARTMENT WAS INTACT AND THERE WAS NO DAMAGE TO THE RETURNED BATTERY CAP. THE BATTERY CAP CONTACT DIMENSIONS WERE WITHIN SPECIFICATIONS. THE RETURNED BATTERY CAP WAS ABLE TO FULLY TIGHTEN ON TO THE PUMP WITH NO POWER LOSS. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO POWER INTERRUPTIONS. THE PUMP COVER WAS OPENED AND NO DAMAGE TO THE POWER CIRCUIT OR EVIDENCE OF INTERNAL MOISTURE WAS FOUND. THE COMPLAINT WAS CONFIRMED IN THE PUMP¿S HISTORY BUT NOT DUPLICATED DURING TESTING. UNRELATED TO THE COMPLAINT, THE DISPLAY SCREEN WAS DISCOLORED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. IT WAS REPORTED THAT THE PUMP INTERMITTENTLY LOST POWER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 309629 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 10 YR |