FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 3831336 · Received May 24, 2014

Report

Report Number
2531779-2014-14711
Event Type
Malfunction
Date Received
May 24, 2014
Report Date
May 16, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE CARTRIDGE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 06/16/2013 -DEVICE EVALUATION: THE CARTRIDGE HAS NOT BEEN RETURNED TO ANIMAS. A RETAIN SAMPLE FROM THE SAME LOT OF CARTRIDGE WAS TESTED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A VISUAL INSPECTION OF THE RETAIN CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A FILL TEST, A FORCE TEST AND A LEAK TEST WERE PERFORMED ON THE RETAIN SAMPLE AND IT MET PASSING CRITERIA. THE CARTRIDGE FORCE READINGS WERE CONFIRMED TO BE WITHIN SPECIFICATIONS; NO FAILURES RELATED TO THE LOSS OF PRIME COMPLAINT WERE NOTED. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION. A REVIEW OF INCOMING INSPECTION RECORD INDICATED THAT THE LOT MET RELEASE CRITERIA.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. REPORTEDLY, THE PUMP GAVE MULTIPLE LOSS OF PRIME ALERTS. IT WAS REPORTED THAT THERE WAS NO SUDDEN CHANGE IN FORCE OR TEMPERATURE AND THE CARTRIDGE CAP WAS SECURED PROPERLY. REVIEW OF CARTRIDGE USE TECHNIQUES INDICATED THAT THE INSTRUCTION FOR USE WAS FOLLOWED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
309687 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION D200053

Patients

Seq Age Sex Outcome Treatment
1 43 YR