FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3831291
·
Received May 23, 2014
Report
- Report Number
- 3004209178-2014-84875
- Event Type
- Malfunction
- Date Received
- May 23, 2014
- Date of Event
- February 22, 2014
- Report Date
- May 20, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST AS A RESULT OF A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE DEVICE WAS RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, SCRATCHED SCREEN, CRACKED RESERVOIR TUBE, SCRATCHED RESERVOIR TUBE WINDOW AND CRACKED BATTERY TUBE THREADS.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED. TROUBLESHOOTING WAS PERFORMED, AND THE CALLER WAS UNSURE IF THE DRIVE SUPPORT CAP WAS PROTRUDED. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 308063 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR |