FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3831117 · Received May 23, 2014

Report

Report Number
1416980-2014-16924
Event Type
Malfunction
Date Received
May 23, 2014
Date of Event
March 18, 2014
Report Date
May 1, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS DETERMINED TO BE A USE ERROR OF SETTING THE TIDAL TOTAL UF REMOVAL TOO LOW. HOMECHOICE APD SYSTEMS TRAINER¿S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. ALL RETURNED HOMECHOICE DEVICES RECEIVE A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 00:26:01. DURING NIGHT DRAIN CYCLE FOUR, THE PATIENT'S ULTRAFILTRATION READING WAS 919ML, INDICATING THE HOME PATIENT (HP) DRAINED 859ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1200ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308128 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 39 YR