FDA Adverse Event Malfunction Summary report: N

UV FLASH TRANSFER SETS

MDR report key: 3831102 · Received May 23, 2014

Report

Report Number
1416980-2014-16910
Event Type
Malfunction
Date Received
May 23, 2014
Date of Event
May 2, 2014
Report Date
May 2, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS REPORTED TO BE AVAILABLE BUT HAS NOT YET BEEN RECEIVED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE TRANSFER SET WAS RETURNED, AND THE EVALUATION WAS COMPLETED. VISUAL INSPECTION (MICROSCOPIC AND NAKED EYE) DISCOVERED DAMAGE TO THE PATIENT ADAPTER. UNDERWATER LEAK, CLEAR PASSAGE AND CLAMP FUNCTIONAL TESTS WERE PERFORMED WITH NO ISSUES NOTED. THE INTERNAL DIAMETER OF THE PATIENT ADAPTER WAS MEASURED, AND IT WAS WITHIN SPECIFICATION LIMITATIONS. A BATCH REVIEW OF THE ASSOCIATED LOT NUMBER WAS PERFORMED, AND IT REVEALED NO ABNORMALITIES DURING THE MANUFACTURE OF THE LOT. ALTHOUGH DAMAGE WAS FOUND ON THE PATIENT ADAPTER, THE TRANSFER SET WAS ABLE TO PASS ALL SUBSEQUENT TESTING AND MEASURED WITHIN SPECIFICATION LIMITS. THE REPORTED CONNECTION ISSUE WAS UNABLE TO BE VERIFIED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A GAP WAS OBSERVED BETWEEN THE PATIENT CONNECTOR OF THE UV FLASH TRANSFER SETS AND THE TITANIUM ADAPTOR. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 2 OF 2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308117 UV FLASH TRANSFER SETS SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME H13J10038

Patients

Seq Age Sex Outcome Treatment
1