HOMECHOICE

MDR report key: 3831064 · Received May 23, 2014

Report

Report Number
1416980-2014-16903
Event Type
Malfunction
Date Received
May 23, 2014
Date of Event
May 2, 2014
Report Date
May 2, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE HOMECHOICE (HC) DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE EVENT HISTORY LOG REVIEW SHOWED NO KEYSTROKES, PROGRAMMING, OR USE RELATED EVENTS THAT INDICATED AND/OR CONTRIBUTED TO THE RITE (RETURNED INSTRUMENT TEST EVALUATION) FAILURE. AN EXTERNAL/INTERNAL INSPECTION WAS PERFORMED WITH NO ISSUES NOTED. TEMPERATURE CONFIRMATION TESTING WAS PERFORMED AND THE RESULTS WERE FOUND TO BE WITHIN SPECIFICATIONS. THE DEVICE WAS DETERMINED TO MEET ELECTRICAL PERFORMANCE SPECIFICATION REQUIREMENTS PER RITE (RETURNED INSTRUMENT TEST EVALUATION) TESTING. THE DEVICE FAILED RITE FUNCTIONAL TESTING AS FLUID WAS TRANSFERRED ABOVE THE ALLOWABLE SPECIFICATION RANGE DURING VOLUMETRIC ACCURACY TESTING. PAL PERFORMED A MORE DETAILED INSPECTION OF THE DOOR ASSEMBLY. THE INSPECTION SHOWED THAT THE PISTON FOAM WAS DETERIORATED. THE RESULTS OF THE EVALUATION REVEALED THE CAUSE OF THE FAILURE TO BE THE DETERIORATED PISTON FOAM. THE PISTON FOAM WAS TO BE SCRAPPED AND THE DEVICE WAS SENT FOR SERVICING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BUT THE EVALUATION HAS YET TO BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED FLUID VOLUME ACCURACY TESTING. DEVICE FAILED DURING EVALUATION, NO PATIENT INVOLVED.