FREESTYLE INSULINX
Report
- Report Number
- 2954323-2014-00581
- Event Type
- Malfunction
- Date Received
- May 23, 2014
- Date of Event
- May 2, 2014
- Report Date
- May 30, 2014
- Product Code
- NBW
- PMA / PMN Number
- K120568
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
THE METER WAS RETURNED AND INVESTIGATED WITH RETURNED TEST STRIPS. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE READINGS THE CUSTOMER REPORTED WERE FOUND IN THE METER MEMORY. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
GIVEN NO TEST STRIPS WERE RETURNED FOR INVESTIGATION, RETAINED TEST STRIP SAMPLES FROM THE SAME LOT REPORTED BY THE CUSTOMER (1374829) WERE TESTED WITH CONTROL SOLUTION. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED. THE COMPLAINT IS NOT CONFIRMED.
THIS IS A CORRECTION REPORT. CORRECTED VERBIAGE IS AS FOLLOWS: THE METER WAS RETURNED AND INVESTIGATED WITH RETAINED TEST STRIPS. THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE READINGS THE CUSTOMER REPORTED WERE FOUND IN THE METER MEMORY. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 487 MG/DL, 254 MG/DL, 139 MG/DL, AND 118 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 307885 | FREESTYLE INSULINX | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 1374829 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 88 YR |