FDA Adverse Event Malfunction Summary report: N

INTERMATE

MDR report key: 3831032 · Received May 23, 2014

Report

Report Number
1416980-2014-16925
Event Type
Malfunction
Date Received
May 23, 2014
Report Date
May 1, 2014
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
PMA / PMN Number
K910425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE SAMPLE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION IDENTIFIED THAT THE BLUE WINGED LUER CAP WAS MISSING, VERIFYING THE REPORTED CONDITION. THE CAUSE OF THE MISSING COMPONENT IS UNKNOWN. A CAPA HAS BEEN OPENED TO ADDRESS THIS ISSUE. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE IS REPORTED TO BE AVAILABLE FOR EVALUATION. IF ANY RELEVANT INFORMATION IS OBTAINED THAT IS RELATED TO THE REPORTED EVENT, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN INTERMATE HAD A MISSING CAP. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
307631 INTERMATE PUMP, INFUSION FRN BAXTER HEALTHCARE - IRVINE 12H014

Patients

Seq Age Sex Outcome Treatment
1