FDA Adverse Event Malfunction Summary report: N

NC TREK CORONARY DILATATION CATHETER

MDR report key: 3830982 · Received May 23, 2014

Report

Report Number
2024168-2014-03322
Event Type
Malfunction
Date Received
May 23, 2014
Date of Event
April 23, 2014
Report Date
May 1, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
LOX
PMA / PMN Number
K110134
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED SHAFT DETACHMENT WAS CONFIRMED. BASED ON ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES OR EXCEPTIONS FOR THIS LOT. A QUERY OF THE ELECTRONIC COMPLAINT HANDLING DATABASE REVEALED NO OTHER SIMILAR INCIDENTS REPORTED FROM THIS LOT. BASED ON THE INFORMATION REVIEWED, THERE IS NO EVIDENCE TO INDICATE THE PRESENCE OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING REMOVAL OF THE NC TREK BALLOON FROM THE PACKAGING, THE PROXIMAL SHAFT BECAME SEPARATED. NO RESISTANCE WAS NOTED DURING REMOVAL OF THE DEVICE FROM THE PACKAGING. THE DEVICE WAS NOT USED ON THE PATIENT. ANOTHER NC TREK BALLOON OF THE SAME SIZE WAS USED ON THE PATIENT. THERE WAS NO CLINICALLY SIGNIFICANT DELAY OF PROCEDURE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
307573 NC TREK CORONARY DILATATION CATHETER CORONARY DILATATION CATHETER LOX AV-TEMECULA-CT 40307G1

Patients

Seq Age Sex Outcome Treatment
1