FDA Adverse Event Malfunction Summary report: N

PROGRASP FORCEPS INSTRUMENT

MDR report key: 3830852 · Received May 23, 2014

Report

Report Number
2955842-2014-03202
Event Type
Malfunction
Date Received
May 23, 2014
Date of Event
May 15, 2014
Report Date
May 21, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS FOUND THE INSTRUMENT'S GRIP CABLE TO BE DERAILED. THE GRIP CABLE WAS DERAILED AT THE DISTAL IDLER PULLEY. YAW MOTION WAS NON-INTUITIVE AS A RESULT. THE OTHER CABLES AT THE WRIST OF THE INSTRUMENT WERE UNDAMAGED. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THE INSTRUMENT'S MAIN TUBE HAD SCRATCH MARKS/ABRASIONS. THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE NOT ALIGNED WITH THE TUBE AXIS. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE VARIOUS SCRATCH MARKS ON THE MAIN TUBE WITH LIGHT MATERIAL REMOVAL FOUND DURING FAILURE ANALYSIS IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT IN CENTRAL PROCESSING, IT WAS OBSERVED THAT THE TIPS ON THE PROGRASP FORCEPS INSTRUMENT WILL NOT GO STRAIGHT. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
307916 PROGRASP FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-10 M10140120 248

Patients

Seq Age Sex Outcome Treatment
1