FDA Adverse Event
Malfunction
Summary report: N
MAESTRO MEDIUM FIXED DURAGUARD
MDR report key: 3830834
·
Received May 23, 2014
Report
- Report Number
- 0001811755-2014-01913
- Event Type
- Malfunction
- Date Received
- May 23, 2014
- Date of Event
- April 26, 2014
- Report Date
- April 26, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HBB
- PMA / PMN Number
- K041754
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED EVENT OF A BENT FOOT ON THE DURAGUARD WAS CONFIRMED BY A STRYKER TECHNICIAN THROUGH VISUAL INSPECTION. BASED ON THE RISK DOCUMENTATION A BENT DURAGUARD CAN OCCUR WHEN EXCESSIVE SIDE LOAD IS APPLIED TO THE FEATURE. DEVICE WAS PLACED IN PARTS RETENTION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DURAGUARD HAS A BENT FOOT. IT WAS FOUND DURING ROUTINE MAINTENANCE. NO ASSOCIATED PROCEDURE, NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 307656 | MAESTRO MEDIUM FIXED DURAGUARD | MOTOR, DRILL, PNEUMATIC | HBB | STRYKER INSTRUMENTS-KALAMAZOO | 12037 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |