FDA Adverse Event Malfunction Summary report: N

EPIX UNIVERSAL CLIP APPLIER

MDR report key: 3830806 · Received May 23, 2014

Report

Report Number
2027111-2014-00261
Event Type
Malfunction
Date Received
May 23, 2014
Date of Event
May 9, 2014
Report Date
August 1, 2014
Manufacturer
APPLIED MEDICAL
Product Code
FZP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: THE EVENT UNIT WAS RETURNED FOR EVALUATION. UPON INSPECTION, ENGINEERING FOUND A DENT IN THE SHAFT OF THE DEVICE NEAR THE KNOB AND CONFIRMED YOUR EXPERIENCE WITH THE DEVICE. THE UNIT WAS DISASSEMBLED FOR FURTHER EVALUATION AND ALL INTERNAL COMPONENTS MET SPECIFICATIONS. THE ROOT CAUSE OF THE JAWS LOCKING IS DUE TO THE BEND IN THE SHAFT OF THE DEVICE, WHICH PREVENTED INTERNAL COMPONENTS FROM MOVING PROPERLY. ALL CLIP APPLIERS UNDERGO 100% VISUAL AND FUNCTIONAL INSPECTION DURING THE MANUFACTURING AND ASSEMBLY PROCESS. AS A PART OF THIS PROCESS, EACH UNIT IS INSPECTED FOR CLIP FEEDING DURING MANUFACTURING PRIOR TO PACKAGING. THIS DOCUMENT REPRESENTS OUR FINAL REPORT.

Additional Manufacturer Narrative · 1

RA HAS JUST RECEIVED THE INCIDENT DEVICE AND HAS BEEN ASSIGNED TO ENGINEERING FOR EVALUATION. A FOLLOW-UP REPORT WILL BE SENT UPON COMPLETION OF INVESTIGATION. IN ACCORDANCE TO 21 CFR 803.56, IF WE OBTAIN ADDITIONAL INFORMATION, WHICH WAS NOT KNOWN OR WAS NOT AVAILABLE WHEN THE INITIAL REPORT WAS SUBMITTED, THEN THE SUPPLEMENTAL REPORT WILL BE SUBMITTED TO THE FDA.

Description of Event or Problem · 1

LAP CHOLE- "AS PER (B)(6), DOCTOR ATTEMPTED TO APPLY CLIP ONTO THE CYSTIC DUCT AND WHEN SHE SQUEEZED THE TRIGGER THE JAWS CLOSED ONTO THE DUCT AND WOULD NOT REOPEN. IT STAYED LOCKED ON THE DUCT. SHE TRIED TO SQUEEZE TRIGGER AGAIN TO RELEASE AND A CLIP FELL OUT OF THE APPLIER." PATIENT STATUS: "GOOD".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
307427 EPIX UNIVERSAL CLIP APPLIER CLIP, IMPLANTABLE, CODE:FZP FZP APPLIED MEDICAL CA500 1214342

Patients

Seq Age Sex Outcome Treatment
1 Other