MAESTRO SMALL STEERING DURAGUA

MDR report key: 3830606 · Received May 23, 2014

Report

Report Number
0001811755-2014-01892
Event Type
Malfunction
Date Received
May 23, 2014
Date of Event
April 29, 2014
Report Date
April 29, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED EVENT OF A BENT FOOT ON THE DURAGUARD WAS CONFIRMED BY A STRYKER SERVICE TECHNICIAN THROUGH VISUAL INSPECTION. BASED ON THE RISK DOCUMENTATION A BENT DURAGUARD CAN OCCUR IF EXCESSIVE SIDE LOAD IS APPLIED TO THE FEATURE. DEVICE IS PLACED IN PARTS RETENTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DURAGAURD HAS A BENT FOOT. IT WAS FOUND DURING SERVICE AT THE MANUFACTURER FACILITY. NO ASSOCIATED PROCEDURE, NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.