FDA Adverse Event Malfunction Summary report: N

ETS LINEAR CUTTER

MDR report key: 3830498 · Received May 23, 2014

Report

Report Number
3005075853-2014-03537
Event Type
Malfunction
Date Received
May 23, 2014
Date of Event
April 23, 2014
Report Date
April 23, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE ATS45 DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH FOUR 6R45B CARTRIDGES PRESENT. THE RELOADS WERE RECEIVED FULLY FIRED. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. EVENT COULD NOT BE CONFIRMED AS THE DEVICE PERFORMED WITHOUT ANY DIFFICULTIES NOTED DURING THE FUNCTIONAL TESTING. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN BARIATRIC PROCEDURE, DURING THE 4TH FIRING OF THE DEVICE, IT STAPLED BUT ONLY CUT HALFWAY ALONG THE STAPLE LINE. NO PATIENT CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308545 ETS LINEAR CUTTER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 RELOAD - 6R45B X 4