FDA Adverse Event Malfunction Summary report: N

FREESTYLE FREEDOM LITE

MDR report key: 3830496 · Received May 23, 2014

Report

Report Number
2954323-2014-00580
Event Type
Malfunction
Date Received
May 23, 2014
Date of Event
April 19, 2014
Report Date
May 1, 2014
Product Code
NBW
PMA / PMN Number
K092602
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 1

REQUESTED PRODUCT WAS NOT RECEIVED FOR INVESTIGATION. RETAINED TEST STRIP SAMPLES FROM THE SAME LOT REPORTED BY THE CUSTOMER (1360041) WERE TESTED WITH CONTROL SOLUTION INSTEAD. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED. THE COMPLAINT IS NOT CONFIRMED.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 300 MG/DL AND 31 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308864 FREESTYLE FREEDOM LITE BLOOD GLUCOSE MONITORING SYSTEM NBW 1360041

Patients

Seq Age Sex Outcome Treatment
1