PWRD ECH FLEX 60MM
Report
- Report Number
- 3005075853-2014-03528
- Event Type
- Malfunction
- Date Received
- May 23, 2014
- Date of Event
- April 30, 2014
- Report Date
- May 1, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K110385
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS FOUND THAT ONE PLE60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH TWO ECR60G CARTRIDGE RELOADS PRESENT AND SIX ECR60D CARTRIDGE RELOADS PRESENT. THE RETURNED RELOADS (C, D, E, F, G, H, I, J) WERE RECEIVED FULLY FIRED. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE STRAIGHT POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. HOWEVER, THE KNIFE WAS NOTED NICKED. ONE POSSIBLE CAUSE FOR THIS TYPE OF DAMAGE TO THE KNIFE IS WHEN THE DEVICE IS FIRED OVER AN ALREADY EXISTING STAPLE LINE, HARD OBJECT OR THICKER TISSUE THAN INDICATED. REPEATEDLY FIRING ACROSS EXISTING STAPLE LINES CAN ALSO REDUCE THE ABILITY TO CUT CLEANLY. TO MITIGATE THE POTENTIAL FOR STAPLES GETTING INTO THE CARTRIDGE AND INTERFERING WITH THE KNIFE PATH DURING DEVICE FIRING, PRIOR TO RELOADING THE DEVICE, RINSE THE ANVIL AND CARTRIDGE JAW IN STERILE SOLUTION AND THEN WIPE THE ANVIL AND CARTRIDGE JAW TO CLEAN ANY FORMED BUT UNUSED STAPLES FROM THE DEVICE. ADDITIONALLY, PROPER CARE SHOULD BE TAKEN WHEN PLACING THE DEVICE ON THE TISSUE TO BE STAPLED, TO ENSURE THAT NO HARD OBSTRUCTION SUCH AS A CLIP IS INCLUDED WITH THE TISSUE INSIDE THE JAWS. AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THAT THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. THE EVENT COULD NOT BE CONFIRMED AS NO INCOMPLETE STAPLE LINE WAS NOTED DURING FUNCTIONAL TESTING. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A ROBOTIC SLEEVE GASTRECTOMY PROCEDURE, THE LAST FIRING OF THE STOMACH WITH GOLD LOAD (ECR60D) THE STAPLES FIRED COMPLETELY ON THE SPECIMEN SIDE AND ON THE PATIENT SIDE THERE WAS A GAP OF NO STAPLES AT THE END OF THE STOMACH. THE SURGEON SEWED THE HOLE CLOSED TO COMPLETE THE CASE. THERE WERE NO PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 307804 | PWRD ECH FLEX 60MM | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | RELOADS-ECR60G ECR60D |