SAPIEN TRANSCATHETER HEART VALVE
Report
- Report Number
- 2015691-2014-01224
- Event Type
- Injury
- Date Received
- May 23, 2014
- Date of Event
- April 28, 2014
- Report Date
- April 28, 2014
- Manufacturer
- EDWARDS LIFESCIENCES
- Product Code
- NPT
- PMA / PMN Number
- P110021
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
PER THE INSTRUCTIONS FOR USE, PARAVALVULAR LEAK (PVL) IS A POTENTIAL ADVERSE EVENT ASSOCIATED WITH BIOPROSTHETIC HEART VALVES AND THE TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVR) PROCEDURE. THE PATIENT SCREENING MANUAL AND THE PROCEDURE DIDACTIC IDENTIFY SEVERAL PROCEDURAL AND ANATOMICAL FACTORS WHICH COULD CONTRIBUTE TO PVL, INCLUDING DEVICE MALPOSITION, INACCURATE MEASUREMENT OF THE NATIVE VALVE ANNULUS, UNEVEN DISTRIBUTION OF CALCIUM ON THE NATIVE VALVE, BULKY OR SEVERE CALCIFICATION, AN ELLIPTICAL ANNULUS SHAPE AND VALVE UNDER-SIZING. PHYSICIANS ARE EXTENSIVELY TRAINED BY EDWARDS BEFORE THEY ARE QUALIFIED TO USE THE SAPIEN THV. THE PATIENT SCREENING MANUAL INSTRUCTS THE OPERATOR ON PROPER AORTIC VALVE & ROOT ASSESSMENT, INCLUDING THE USE OF ECHO, AORTOGRAM AND CT TO APPROPRIATELY MEASURE THE ANNULUS DIAMETER, CONTENT AND DISTRIBUTION OF CALCIUM, AND LEAFLET CHARACTERISTICS. CONTRAINDICATIONS, IMPORTANT CONSIDERATIONS WHEN ASSESSING THE VALVE, AND CHOOSING THE PROPER THV ARE ALSO DISCUSSED. THE THV TRAINING MANUALS ALSO INSTRUCT THE OPERATOR ON PROPER POSITIONING AND DEPLOYMENT OF THE VALVE, INCLUDING ALL PROCEDURAL AND ANATOMICAL CONSIDERATIONS. DEVICE PREPARATION, APPROACH, DEPLOYMENT, IMAGING, PROCEDURE-SPECIFIC TRAINING MANUALS AND PROCTORED PROCEDURES ARE ALSO INCLUDED. IN THIS CASE, THE CAUSE OF THE PVL COULD NOT BE CONFIRMED; HOWEVER PATIENT FACTORS (ELLIPTICAL NATIVE ANNULUS) COULD HAVE CONTRIBUTED TO THE PARAVALVULAR LEAK. THE IFU AND TRAINING MANUALS HAVE BEEN REVIEWED AND NO INADEQUACIES HAVE BEEN IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A (B)(4) BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS (B)(4) REVIEW. NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED.
POST DEPLOYMENT OF A SAPIEN VALVE, MODERATE TO SEVERE PARAVALVULAR LEAK (PVL) WAS NOTED ON ECHO. A SECOND INFLATION OF THE DELIVERY BALLOON WAS PERFORMED AND THE VALVE WAS REASSESSED; HOWEVER THE PVL REMAINED UNCHANGED. THE SURGICAL TEAM DECIDED TO IMPLANT A SECOND VALVE. THE SECOND VALVE WAS DEPLOYED AND TRACE TO MILD PVL WAS ASSESSED BY ECHO. THE TEAM DECIDED THAT SINCE THERE WAS ONLY TRACE TO MILD PVL THERE WAS NO NEED FOR ADDITIONAL TREATMENT. THE DELIVERY SYSTEM WAS REMOVED. THE PATIENT¿S NATIVE ANNULUS WAS MODERATELY CALCIFIED AND MEASURED 19MM BY 24MM ON CT, WITH AN AREA OF 384.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 307658 | SAPIEN TRANSCATHETER HEART VALVE | AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED | NPT | EDWARDS LIFESCIENCES | 9000TFX23 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Required Intervention |