FDA Adverse Event Malfunction Summary report: N

ADVANCE BED

MDR report key: 3830181 · Received May 12, 2014

Report

Report Number
1824206-2014-01467
Event Type
Malfunction
Date Received
May 12, 2014
Date of Event
April 14, 2014
Report Date
April 14, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE COMMUNICATION HOUSING ASSEMBLY IS NOT FUNCTIONING DUE TO NORMAL WEAR AND TEAR. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE SIDE RAIL MOUNT BRACKET TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE ACCOUNT REPORTED THE BED EXIT WAS NOT ALARMING. THE BED WAS LOCATED IN (B)(6) AT THE FACILITY. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
283344 ADVANCE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1135

Patients

Seq Age Sex Outcome Treatment