REPLY
Report
- Report Number
- 1000165971-2014-00298
- Event Type
- Injury
- Date Received
- May 23, 2014
- Date of Event
- April 8, 2014
- Report Date
- April 22, 2014
- Manufacturer
- SORIN GROUP ITALIA S.R.L. - CRM FACILITY
- Product Code
- NVZ
- PMA / PMN Number
- PP950029
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
INVESTIGATION RESULTS OF THE RETURNED UNIT CONFIRMED THAT THE END OF LIFE WAS REACHED; HOWEVER CURRENT CONSUMPTION OF THE ELECTRONIC MODULE WAS NORMAL.
REPORTEDLY, DURING THE FOLLOW-UP OF THE SUBJECT PACEMAKER PERFORMED ON (B)(6) 2014, END OF LIFE (EOL) WAS REACHED. THE DEVICE WAS EXPLANTED AND RETURNED FOR ANALYSIS. IT WAS REPORTED ALSO THAT DURING PREVIOUS FOLLOW-UP PERFORMED ON (B)(6) 2013, THE BATTERY IMPEDANCE WAS AT 2.8 KOHM AND IT WAS DISPLAYED A TIME TO ELECTIVE REPLACEMENT INDICATOR (ERI) OF 13 MONTHS.
REPORTEDLY, DURING THE FOLLOW-UP OF THE SUBJECT PACEMAKER PERFORMED ON(B)(6) 2014, END OF LIFE (EOL) WAS REACHED. THE DEVICE WAS EXPLANTED AND RETURNED FOR ANALYSIS. IT WAS REPORTED ALSO THAT DURING PREVIOUS FOLLOW-UP PERFORMED ON (B)(6) 2013, THE BATTERY IMPEDANCE WAS AT 2.8 KOHM AND IT WAS DISPLAYED A TIME TO ELECTIVE REPLACEMENT INDICATOR (ERI) OF 13 MONTHS.
REPORTEDLY, DURING THE FOLLOW-UP OF THE SUBJECT PACEMAKER PERFORMED ON (B)(6) 2014, END OF LIFE (EOL) WAS REACHED. THE DEVICE WAS EXPLANTED AND RETURNED FOR ANALYSIS. IT WAS REPORTED ALSO THAT DURING PREVIOUS FOLLOW-UP PERFORMED ON (B)(6) 2013, THE BATTERY IMPEDANCE WAS AT 2.8 KOHM AND IT WAS DISPLAYED A TIME TO ELECTIVE REPLACEMENT INDICATOR (ERI) OF 13 MONTHS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 308514 | REPLY | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | SORIN GROUP ITALIA S.R.L. - CRM FACILITY | ESPRIT DR | 2420 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |