OVATIO
Report
- Report Number
- 1000165971-2014-00297
- Event Type
- Injury
- Date Received
- May 23, 2014
- Date of Event
- April 21, 2014
- Report Date
- April 21, 2014
- Manufacturer
- SORIN CRM
- Product Code
- MRM
- PMA / PMN Number
- PP980049
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PO
- Reporter Occupation
- OTHER
Narratives
PRELIMINARY ANALYSIS OF THE RETURNED DEVICE CONFIRMED IT OPERATED AS SPECIFIED.
REPORTEDLY, DURING THE INTERROGATION OF THE SUBJECT ICD, THE ELECTIVE REPLACEMENT INDICATOR (ERI) WAS REACHED. THE PHYSICIAN WANTED TO PERFORM A CARDIOVERSION TO FREE THE PATIENT FROM AN ATRIAL FIBRILLATION. A MANUAL 34J-SHOCK WAS ATTEMPTED BUT AFTER A FEW SECONDS, NO SHOCK WAS DELIVERED AND THE "CHARGING" MESSAGE WAS SUBSTITUTED FOR "WAITING...". NOTHING ELSE HAPPENED. THE PATIENT WAS THEN PROVIDED WITH A EXTERNAL SHOCK THAT CONVERTED THE RHYTHM TO SINUSAL RHYTHM. THE PHYSICIAN DECIDED TO REPLACE THE DEVICE ON (B)(6) 2014 AND WILL BE RETURNED FOR ANALYSIS.
REPORTEDLY, DURING THE INTERROGATION OF THE SUBJECT ICD, THE ELECTIVE REPLACEMENT INDICATOR (ERI) WAS REACHED. THE PHYSICIAN WANTED TO PERFORM A CARDIOVERSION TO FREE THE PATIENT FROM AN ATRIAL FIBRILLATION. A MANUAL 34J-SHOCK WAS ATTEMPTED BUT AFTER A FEW SECONDS, NO SHOCK WAS DELIVERED AND THE "CHARGING" MESSAGE WAS SUBSTITUTED FOR "WAITING...". NOTHING ELSE HAPPENED. THE PATIENT WAS THEN PROVIDED WITH A EXTERNAL SHOCK THAT CONVERTED THE RHYTHM TO SINUSAL RHYTHM. THE PHYSICIAN DECIDED TO REPLACE THE DEVICE ON (B)(6) 2014 AND WILL BE RETURNED FOR ANALYSIS.
REPORTEDLY, DURING THE INTERROGATION OF THE SUBJECT ICD, THE ELECTIVE REPLACEMENT INDICATOR (ERI) WAS REACHED. THE PHYSICIAN WANTED TO PERFORM A CARDIOVERSION TO FREE THE PATIENT FROM AN ATRIAL FIBRILLATION. A MANUAL 34J-SHOCK WAS ATTEMPTED BUT AFTER A FEW SECONDS, NO SHOCK WAS DELIVERED AND THE "CHARGING" MESSAGE WAS SUBSTITUTED FOR "WAITING..." NOTHING ELSE HAPPENED. THE PATIENT WAS THEN PROVIDED WITH A EXTERNAL SHOCK THAT CONVERTED THE RHYTHM TO SINUSAL RHYTHM. THE PHYSICIAN DECIDED TO REPLACE THE DEVICE ON 23 APRIL 2014 AND WILL BE RETURNED FOR ANALYSIS.
REPORTEDLY, DURING THE INTERROGATION OF THE SUBJECT ICD, THE ELECTIVE REPLACEMENT INDICATOR (ERI) WAS REACHED. THE PHYSICIAN WANTED TO PERFORM A CARDIOVERSION TO FREE THE PATIENT FROM AN ATRIAL FIBRILLATION. A MANUAL 34J-SHOCK WAS ATTEMPTED BUT AFTER A FEW SECONDS, NO SHOCK WAS DELIVERED AND THE "CHARGING" MESSAGE WAS SUBSTITUTED FOR "WAITING...". NOTHING ELSE HAPPENED. THE PATIENT WAS THEN PROVIDED WITH A EXTERNAL SHOCK THAT CONVERTED THE RHYTHM TO SINUSAL RHYTHM. THE PHYSICIAN DECIDED TO REPLACE THE DEVICE ON (B)(6) 2014 AND WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 308328 | OVATIO | DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER | MRM | SORIN CRM | OVATIO DR 6550 | S060223 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |