FDA Adverse Event Injury Summary report: N

OVATIO

MDR report key: 3830095 · Received May 23, 2014

Report

Report Number
1000165971-2014-00297
Event Type
Injury
Date Received
May 23, 2014
Date of Event
April 21, 2014
Report Date
April 21, 2014
Manufacturer
SORIN CRM
Product Code
MRM
PMA / PMN Number
PP980049
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRELIMINARY ANALYSIS OF THE RETURNED DEVICE CONFIRMED IT OPERATED AS SPECIFIED.

Description of Event or Problem · 1

REPORTEDLY, DURING THE INTERROGATION OF THE SUBJECT ICD, THE ELECTIVE REPLACEMENT INDICATOR (ERI) WAS REACHED. THE PHYSICIAN WANTED TO PERFORM A CARDIOVERSION TO FREE THE PATIENT FROM AN ATRIAL FIBRILLATION. A MANUAL 34J-SHOCK WAS ATTEMPTED BUT AFTER A FEW SECONDS, NO SHOCK WAS DELIVERED AND THE "CHARGING" MESSAGE WAS SUBSTITUTED FOR "WAITING...". NOTHING ELSE HAPPENED. THE PATIENT WAS THEN PROVIDED WITH A EXTERNAL SHOCK THAT CONVERTED THE RHYTHM TO SINUSAL RHYTHM. THE PHYSICIAN DECIDED TO REPLACE THE DEVICE ON (B)(6) 2014 AND WILL BE RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

REPORTEDLY, DURING THE INTERROGATION OF THE SUBJECT ICD, THE ELECTIVE REPLACEMENT INDICATOR (ERI) WAS REACHED. THE PHYSICIAN WANTED TO PERFORM A CARDIOVERSION TO FREE THE PATIENT FROM AN ATRIAL FIBRILLATION. A MANUAL 34J-SHOCK WAS ATTEMPTED BUT AFTER A FEW SECONDS, NO SHOCK WAS DELIVERED AND THE "CHARGING" MESSAGE WAS SUBSTITUTED FOR "WAITING...". NOTHING ELSE HAPPENED. THE PATIENT WAS THEN PROVIDED WITH A EXTERNAL SHOCK THAT CONVERTED THE RHYTHM TO SINUSAL RHYTHM. THE PHYSICIAN DECIDED TO REPLACE THE DEVICE ON (B)(6) 2014 AND WILL BE RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

REPORTEDLY, DURING THE INTERROGATION OF THE SUBJECT ICD, THE ELECTIVE REPLACEMENT INDICATOR (ERI) WAS REACHED. THE PHYSICIAN WANTED TO PERFORM A CARDIOVERSION TO FREE THE PATIENT FROM AN ATRIAL FIBRILLATION. A MANUAL 34J-SHOCK WAS ATTEMPTED BUT AFTER A FEW SECONDS, NO SHOCK WAS DELIVERED AND THE "CHARGING" MESSAGE WAS SUBSTITUTED FOR "WAITING..." NOTHING ELSE HAPPENED. THE PATIENT WAS THEN PROVIDED WITH A EXTERNAL SHOCK THAT CONVERTED THE RHYTHM TO SINUSAL RHYTHM. THE PHYSICIAN DECIDED TO REPLACE THE DEVICE ON 23 APRIL 2014 AND WILL BE RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

REPORTEDLY, DURING THE INTERROGATION OF THE SUBJECT ICD, THE ELECTIVE REPLACEMENT INDICATOR (ERI) WAS REACHED. THE PHYSICIAN WANTED TO PERFORM A CARDIOVERSION TO FREE THE PATIENT FROM AN ATRIAL FIBRILLATION. A MANUAL 34J-SHOCK WAS ATTEMPTED BUT AFTER A FEW SECONDS, NO SHOCK WAS DELIVERED AND THE "CHARGING" MESSAGE WAS SUBSTITUTED FOR "WAITING...". NOTHING ELSE HAPPENED. THE PATIENT WAS THEN PROVIDED WITH A EXTERNAL SHOCK THAT CONVERTED THE RHYTHM TO SINUSAL RHYTHM. THE PHYSICIAN DECIDED TO REPLACE THE DEVICE ON (B)(6) 2014 AND WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308328 OVATIO DEFIBRILLATOR, IMPLANTABLE, DUAL-CHAMBER MRM SORIN CRM OVATIO DR 6550 S060223

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention