FDA Adverse Event Summary report: N

RAPICIDE GLUTERALDEHYDE

MDR report key: 3830057 · Received May 23, 2014

Report

Report Number
2150060-2014-00018
Date Received
May 23, 2014
Date of Event
May 5, 2014
Report Date
May 23, 2014
Manufacturer
MEDIVATORS
Product Code
MED
PMA / PMN Number
K993042
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MEDIVATORS FIELD SERVICE ENGINEER (FSE) WENT TO FACILITY TO REPAIR THEIR ADVANTAGE PLUS AER AND DISCOVERED RAPICIDE GLUTARALDEHYDE WAS BEING USED AT ROOM TEMPERATURE MANUALLY SOAKING ENDOSCOPES DUE TO THEIR MACHINE BEING DOWN. RAPICIDE IS NOT APPROVED FOR ROOM TEMP MANUAL HIGH LEVEL DISINFECTION ACCORDING TO MEDIVATORS RAPICIDE IFU. THE FIELD SERVICE ENGINEER OBSERVED THAT THE RAPICIDE BOTTLE WERE LABELED (HAND WRITTEN) "USE AS CIDEX." THEY ARE REFERENCING CIDEX OPA MANUAL USE INSTRUCTIONS FOR HIGH LEVEL DISINFECTION INSTRUCTIONS. THE MEDIVATORS FSE INFORMED THE FACILITY THAT THEIR PROCEDURES OF MANUAL DISINFECTION USING THE RAPICIDE WERE INCORRECT. RAPICIDE REQUIRES HEAT TO REACH HIGH LEVEL DISINFECTION FOR ENDOSCOPES. THEREFORE, INADEQUATE AND INAPPRPRIATE REPROCESSING MAY CAUSE PATIENT INJURY VIA CROSS CONTAMINATION. THEIR MEDIVATORS ADVANTAGE PLUS MACHINE IS NOW OPERATING ACCORDING TO SPECIFICATIONS AFTER FSE REPAIRED THE MACHINE. IT WAS REPORTED THE MACHINE WAS DOWN AND THE FACILITY WAS MANUAL REPROCESSING SCOPES FOR 2 DAYS. THIS COMPLAINT WILL CONTINUE TO BE MONITORED WITHIN MEDIVATORS COMPLAINT SYSTEM.

Description of Event or Problem · 1

THE CASE STATES THAT A FACILITY WAS USING RAPICIDE GLUTARALDEHYDE FOR MANUAL CLEANING OF THEIR ENDOSCOPES WHILE THEIR MEDIVATORS ADVANTAGE PLUS MACHINE WAS BEING REPAIRED. MEDIVATORS RAPICIDE IS NOT APPROVED FOR ROOM TEMP MANUAL HLD AS THIS PRODUCT REQUIRES HEAT. REPORTED THE MACHINE HAD BEEN DOWN FOR TWO DAYS. POTENTIAL CROSS CONTAMINATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308751 RAPICIDE GLUTERALDEHYDE HIGH LEVEL DISINFECTANT MED MEDIVATORS ML02-0059

Patients

Seq Age Sex Outcome Treatment
1