AESTIVEA 7900
Report
- Report Number
- 2112667-2014-00037
- Event Type
- Malfunction
- Date Received
- March 11, 2014
- Date of Event
- January 14, 2014
- Report Date
- January 14, 2014
- Manufacturer
- DATEX-OHMEDA
- Product Code
- CBK
- PMA / PMN Number
- K023366
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THIS MALFUNCTION WAS DETERMINED TO BE REPORTABLE AS THIS SAME MALFUNCTION HAS PREVIOUSLY CONTRIBUTED TO A SERIOUS INJURY WITHIN THE LAST TWO YEARS. REFERENCE MDR 2112667-2013-00005. A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE EQUIPMENT AND NOTED THAT THE EXPIRATORY FLOW SENSOR DIAPHRAGM WAS STUCK OPEN. MANUAL MODE OF VENTILATION IS AVAILABLE TO MAINTAIN VENTILATION OF THE PATIENT. FLOW SENSORS OF THIS TYPE ARE CUSTOMER REPLACEABLE, ARE RECOMMENDED FOR REPLACEMENT AFTER 3 MONTHS, AND ARE WARRANTED FOR 6 MONTHS. THE MAINTENANCE SCHEDULE IN THE USER REFERENCE MANUAL STATES: "REPLACE THE DISPOSABLE FLOW SENSOR (PLASTIC). UNDER TYPICAL USE, THE SENSOR MEETS SPECIFICATIONS FOR A MINIMUM OF 3 MONTHS." IN ENGINEERING EVALUATION, THE STUCK DIAPHRAGM HAS BEEN ABLE TO BE REPRODUCED BY: A HARD IMPACT, SUCH AS DROPPING THE FLOW SENSOR, OR BY STICKING AN OBJECT INTO THE FLOW SENSOR, CAUSING THE DIAPHRAGM TO STICK OPEN. IF A SENSOR IS SUBJECTED TO A HARD IMPACT, IT IS STILL UNLIKELY THAT THE DIAPHRAGM WILL GET STUCK IN THE OPEN POSITION. THIS FAILURE MODE REQUIRES AN IMPACT IN A VERY LIMITED ORIENTATION TO RESULT IN THE INERTIA NEEDED TO FORCE THE DIAPHRAGM INTO THE STUCK OPEN POSITION.
DURING A PREOPERATIVE CHECKOUT OF THE EQUIPMENT, THE HOSPITAL NOTED EXPIRATORY FLOW WAS GREATER THAN INSPIRATORY FLOW. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 145978 | AESTIVEA 7900 | ANESTHESIA MACHINE | CBK | DATEX-OHMEDA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |