NA
Report
- Report Number
- 0001056128-2014-00060
- Event Type
- Malfunction
- Date Received
- May 22, 2014
- Date of Event
- January 6, 2014
- Report Date
- April 28, 2014
- Manufacturer
- STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
- Product Code
- NUJ
- PMA / PMN Number
- K111600
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
AN INVESTIGATION HAS NOT BEEN COMPLETED FOR THIS EVENT. ONCE AN INVESTIGATION HAS BEEN COMPLETED A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THE COMPLAINT DEVICE WAS RETURNED TO STRYKER SUSTAINABILITY SOLUTIONS (SSS) FOR EVALUATION. THE DEVICE WAS RECEIVED WITH VISIBLE BIOLOGICAL MATERIAL ON THE DISTAL TIP; EVIDENCE THAT IT WAS CLINICALLY USED. THE JAW FUNCTIONALITY WAS TESTED AND CONFIRMED TO BE ACCEPTABLE AS THE DEVICE WAS ABLE TO ACTUATE, LOCKED/UNLOCK MULTIPLE TIMES. WHILE THE JAWS WERE IN THE LOCKED POSITION, THEY WERE INSPECTED AND FOUND TO BE PROPERLY ALIGNED. THE BLADE TRIGGER WAS TESTED AND VERIFIED TO MAINTAIN PROPER MOVEMENT AND DID NOT STICK OR DERAIL WHEN ACTIVATED. THE DEVICE WAS THEN CONFIRMED TO PASS CUT TESTING AS IT WAS ABLE TO SUCCESSFULLY MAKE THREE (3) CONSECUTIVE CUTS WITHOUT ANY BUNCHING; FRAYING; OR MAKING ANY INCOMPLETE OR UNCLEAN CUTS. THE RESULTS OF THE INVESTIGATION PERFORMED INDICATED THAT THE JAWS OF THE RETURNED DEVICE FUNCTIONED AS INTENDED AS IT WAS ABLE TO ACTUATE (LOCK/UNLOCK) AS INTENDED AND PASSED CUT TESTING. POTENTIAL CAUSES FOR THE REPORTED ISSUE CAN BE ATTRIBUTED BUT NOT LIMITED TO; ESCHAR BUILDUP INSIDE THE GUIDING SLOTS AND JAWS AS IT CAN INHIBIT THE GRASPING/CUTTING ABILITY AND POTENTIALLY CAUSE THE BLADE TO STICK/DRAG CAUSING THE BLADE TO DERAIL AND TEMPORARILY LOCK AS A RESULT OF OVER FILLING THE JAWS WITH TISSUE, AND THEN SUBSEQUENTLY REALIGNING WHILE MANIPULATING THE DEVICE IN THE PATIENT; OR UNKNOWN USER TECHNIQUES WHICH CAUSED THE DEVICE NOT TO CUT. THE INSTRUCTIONS FOR USE (IFU) REPROCESSED HAND ACTIVATED SEALER/DIVIDER STATES: ¿KEEP THE INSTRUMENT JAWS CLEAN. BUILD-UP OF ESCHAR MAY REDUCE SEALING AND/OR CUTTING EFFECTIVENESS. WIPE JAW SURFACES AND EDGES WITH A WET GAUZE PAD AS NEEDED.¿ ¿DO NOT USE THIS DEVICE ON VESSELS IN EXCESS OF 7MM IN DIAMETER.¿ ¿TO ENSURE PROPER FUNCTION, ELIMINATE TENSION ON THE TISSUE WHILE SEALING AND CUTTING.¿ ¿AVOID OVERFILLING THE INSTRUMENT JAWS WITH TISSUE. THIS MAY DAMAGE THE CUTTING MECHANISM OR CAUSE THE BLADE TO DEPLOY OUTSIDE OF ITS GUIDING FEATURES, POSSIBLY RESULTING IN DIFFICULTY OPENING THE JAWS OR UNINTENDED INJURY TO THE USER OR PATIENT.¿ ¿VISUALLY CONFIRM THAT THE JAWS HAVE REACHED THE CLOSED POSITION PRIOR TO ACTIVATING THE CUTTER. FAILURE TO DO SO MAY DAMAGE THE CUTTING MECHANISM OR CAUSE THE BLADE TO DEPLOY OUTSIDE OF ITS GUIDING FEATURES, POSSIBLY RESULTING IN DIFFICULTY OPENING THE JAWS OR UNINTENDED INJURY TO THE USER OR PATIENT.¿ ¿CLOSE THE WHITE MOVABLE HANDLE UNTIL IT CLICKS AND LATCHES IN PLACE.¿ ¿OPEN THE JAWS BY SQUEEZING THE WHITE MOVABLE HANDLE UNTIL IT UNLOCKS, THEN PUSH IT COMPLETELY FORWARD.¿ ¿PRIOR TO ACTIVATING THE CUTTER, CONFIRM THAT THE JAWS ARE IN THE CLOSED POSITION. SPACING BETWEEN JAWS MUST BE LESS THAN TWO MILLIMETERS.¿ THE DEVICE HISTORY RECORDS (DHR) FOR THE COMPLAINT DEVICE INDICATES THAT THE DEVICE PASSED ALL APPLICABLE INSPECTIONS AND TESTS PRIOR TO RELEASE. THIS REPORT WAS FILED DUE TO THE REPORTED EVENT BEING AN INHERENTLY DANGEROUS SITUATION. THE REPORTED EVENT IS NOT OCCURRING MORE FREQUENTLY OR WITH GREATER SEVERITY THAN IS USUAL FOR THE DEVICE.
IT WAS REPORTED THAT DURING A PROCEDURE THE LIGASURE "BLADES ARE CLOSED TOGETHER, WOULD NOT ACTIVATE AND WOULD NOT COAGULATE." THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO MEDICAL TREATMENT OR INTERVENTION REQUIRED AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.
IT WAS REPORTED THAT DURING A PROCEDURE THE LIGASURE "BLADES ARE CLOSED TOGETHER, WOULD NOT ACTIVATE AND WOULD NOT COAGULATE." THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO MEDICAL TREATMENT OR INTERVENTION REQUIRED AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 305782 | NA | ELECTROSURGICAL CUTTING AND COAGULATION DEVICE AND ACCESSORIES | NUJ | STRYKER SUSTAINABILITY SOLUTIONS LAKELAND | LF4200 | 2843430 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |