GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2014-00399
- Event Type
- Malfunction
- Date Received
- May 21, 2014
- Date of Event
- May 15, 2014
- Report Date
- May 15, 2014
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
- Product Code
- GEX
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE GLASS CAP EXHIBITS MODERATE DEVITRIFICATION AT OUTPUT WINDOW; THE METAL CAP EXHIBITS MODERATE DETRITUS ADHESION. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER.
(B)(4) REFERS TO FORWARD FIRING OF THE SIDE FIRING SURGICAL FIBER; A CODE REQUEST HAS BEEN SUBMITTED.
IT WAS REPORTED THAT DURING A PROSTATE PROCEDURE, FORWARD FIRING WAS OBSERVED @ 32,079 JOULES, 5:59 MINUTES OF LASING TIME. NO COMPLICATIONS WERE NOTED, SURGERY WAS COMPLETED WITH A SECOND FIBER WITHOUT INCIDENT AND PATIENT DOING FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 303177 | GREENLIGHT MOXY FIBER OPTIC | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) | 0010-2400 | 344A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |