LIFEVEST WCD 4000 SYSTEM
MDR report key: 3823814
·
Received May 2, 2014
Report
- Report Number
- 3008642652-2014-01349
- Event Type
- Malfunction
- Date Received
- May 2, 2014
- Date of Event
- March 27, 2014
- Report Date
- April 30, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T POWER UP) HAS BEEN CONFIRMED. UPON INVESTIGATION THE MONITOR WAS UNABLE TO POWER UP. THE CAUSE FOR THE FAILURE WAS ISOLATED TO CORROSION ON CONNECTOR J502 ON THE C/A BOARD, CAUSING THE +5V AND -5V POWER SUPPLIES AND CAPACITOR C19 TO SHORT. THE ROOT CAUSE FOR THE CORROSION COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY INGRESS OF AN UNKNOWN LIQUID. NO ADVERSE EVENT RESULTED FROM THE CORROSION. THE PATIENT RECEIVED A REPLACEMENT MONITOR.
Description of Event or Problem · 1
THE MOTHER OF A (B)(6) MALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PATIENT'S MONITOR WAS NOT POWERING UP PROPERLY. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 265260 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |