ZIMMER SKIN GRAFT MESHER
Report
- Report Number
- 1526350-2014-00102
- Event Type
- Malfunction
- Date Received
- February 20, 2014
- Date of Event
- January 1, 2014
- Report Date
- January 22, 2014
- Manufacturer
- ZIMMER SURGICAL
- Product Code
- GFD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- UNKNOWN
Narratives
THE DEVICE WAS RETURNED TO THE MFR FOR REPAIR AND EVALUATION. THE SERVICE RECORD INDICATES THAT THE DEVICE WAS MANUFACTURED ON 08/23/2010 AND WAS LAST REPAIRED ON 04/10/2013 FOR RELATED ISSUE. EVALUATION OF THE DEVICE OBSERVED WORN GEARS, AS EXCESSIVE GALLING OF THE GEAR TEETH OF THE ROLLER GEAR AND BOTH CUTTER GEARS WAS OBSERVED. THE, THE COMB WAS OBSERVED TO BE DAMAGED. PRIOR TO REPAIR, TEST MESH AND ADDITIONALLY CHECK COULD NOT BE PERFORMED DUE TO THE DAMAGED GEARS. REPAIR OF THE DEVICE INCLUDED REPLACEMENT OF THE ROLLER AND ROLLER GEAR, COMB, BOTH SIDE PLATES, AND CARRIER GUIDE. THE CUSTOMER RETURNED TWO CUTTERS FOR EVALUATION. THE GEARS AND BUSHINGS WERE REPLACED TO BOTH CUTTERS, WHICH PASSED EVALUATION. IT WAS NOTED THAT THE CUSTOMER IS A CADAVER TISSUE BANK, WHICH EXPERIENCES HEAVY USAGE OF DEVICES. IMPROPER HANDLING BY THE USER MOST LIKELY CAUSED DAMAGE TO THE ROLLER GEAR, BOTH CUTTER GEARS AND COMB, WHICH MOST LIKELY CAUSED THE CUSTOMERS REPORTED EVENT. THE DEVICE WAS SERVICES AND RETURNED TO THE CUSTOMER.
IT WAS REPORTED THAT THE GEARS WERE DAMAGED ON THE ZIMMER SKIN GRAFT MESHER AND WAS HARD TO CRANK. IT WAS NOTED THAT THE 1.5:1 CUTTER, SERIAL NUMBER (B)(4) WAS NOT MESHING. THE FACILITY REPORTING THE EVENT IS A CADAVER TISSUE BANK. AS SUCH, THERE WAS NO REPORT OR PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 106579 | ZIMMER SKIN GRAFT MESHER | ZIMMER SKIN GRAFT MESHER | GFD | ZIMMER SURGICAL | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ZIMMER CUTTER: 00-7703-015-00, SN (B)(4)| ZIMMER CUTTER: 00-7703-020-00, SN (B)(4) |