FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3823782 · Received May 2, 2014

Report

Report Number
3008642652-2014-01289
Event Type
Malfunction
Date Received
May 2, 2014
Date of Event
April 4, 2014
Report Date
April 23, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 204) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT WAS UNABLE TO COMMUNICATE WITH A MONITOR. THE CAUSE FOR THE FAILURE WAS ISOLATED TO A STRAINED TRUNK CABLE. THE TRUNK CABLE WAS PULLED FROM THE STRAIN RELIEF, DAMAGING CONNECTOR J702 ON THE DISTRIBUTION NODE PCA. THE ROOT CAUSE FOR THE STRAINED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE STRAINED CABLE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

THE SON OF A (B)(6) MALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PATIENT WAS RECEIVING A SERVICE CODE 204. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
264185 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 N/A

Patients

Seq Age Sex Outcome Treatment
1 58 YR