FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3823572
·
Received February 19, 2014
Report
- Report Number
- 1720753-2014-01646
- Event Type
- Malfunction
- Date Received
- February 19, 2014
- Date of Event
- February 2, 2014
- Report Date
- February 19, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE INTERCONNECT CABLE WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM DISPLAYED AN INTERMITTENT COMMUNICATION ERROR MESSAGE. THIS ERROR MAY RESULT IN A SYSTEM LOCK UP, NO BOOT, OR SHUT DOWN SITUATION. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 104292 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |