FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 3823495 · Received February 19, 2014

Report

Report Number
1720753-2014-01666
Event Type
Malfunction
Date Received
February 19, 2014
Date of Event
February 7, 2014
Report Date
February 19, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE PS1 POWER SUPPLY WAS EVALUATED. THE FFB PCB ASSEMBLY WAND INTERCONNECT CABLE WERE ALSO EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY FAILED TO BOOT TO A USABLE STATE AND EXHIBITED INTERMITTENT FUNCTIONALITY. NO PT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104759 8800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 8800

Patients

Seq Age Sex Outcome Treatment
1