FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3823438
·
Received May 1, 2014
Report
- Report Number
- 3008642652-2014-01208
- Event Type
- Malfunction
- Date Received
- May 1, 2014
- Date of Event
- April 4, 2014
- Report Date
- April 17, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. AS RECEIVED, THE CHARGER/ MODEM WAS RESETTING. UPON EVALUATION, THERE WAS CORRUPTED SOFTWARE ON THE FLASH MEMORY. THE CAUSE OF THE RESETS IS THE CORRUPTED SOFTWARE THE ROOT CAUSE OF THE CORRUPTED SOFTWARE CANNOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE CORRUPTED SOFTWARE THE LAST PATIENT TO USE THIS CHARTER/MODEM DID NOT REPORT ANY DEFICIENCIES.
Description of Event or Problem · 1
REVIEW OF SERVICE DATE DETECTED A REPORTABLE PROBLEM. DURING SERVICING, CHARGER/MODEM SN (B)(4) WAS RESETTING. THE LAST PATIENT TO USE THIS CHARGER/ MODEM DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 261947 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |