FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3823438 · Received May 1, 2014

Report

Report Number
3008642652-2014-01208
Event Type
Malfunction
Date Received
May 1, 2014
Date of Event
April 4, 2014
Report Date
April 17, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. AS RECEIVED, THE CHARGER/ MODEM WAS RESETTING. UPON EVALUATION, THERE WAS CORRUPTED SOFTWARE ON THE FLASH MEMORY. THE CAUSE OF THE RESETS IS THE CORRUPTED SOFTWARE THE ROOT CAUSE OF THE CORRUPTED SOFTWARE CANNOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE CORRUPTED SOFTWARE THE LAST PATIENT TO USE THIS CHARTER/MODEM DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

REVIEW OF SERVICE DATE DETECTED A REPORTABLE PROBLEM. DURING SERVICING, CHARGER/MODEM SN (B)(4) WAS RESETTING. THE LAST PATIENT TO USE THIS CHARGER/ MODEM DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
261947 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 N/A

Patients

Seq Age Sex Outcome Treatment
1 NA