PLUM A+ DRIVER ED 2
Report
- Report Number
- 9615050-2014-02209
- Event Type
- Malfunction
- Date Received
- March 20, 2014
- Date of Event
- February 28, 2014
- Report Date
- March 1, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K042081
- Removal / Correction Number
- Z-0624-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS BROKEN. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THE DEVICE DOOR ROLLER WAS BROKEN. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR A REPORT OF BROKEN DOOR ROLLER. NO SPECIFIC INFORMATION REGARDING THE PATIENT, PUMP PROGRAMMING, OR EVENT DETAILS WERE AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING THERAPIES AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS BROKEN. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 166595 | PLUM A+ DRIVER ED 2 | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | PLUM A+ SOFTWARE MODULE, LIST #UNK, SN UNK |