Report Number
1720753-2014-02700
Event Type
Malfunction
Date Received
March 26, 2014
Date of Event
March 4, 2014
Report Date
March 26, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE CONTROL PANEL ENCODER PCB AND PS2 WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE FSE REPORTED INTERMITTENT COMMUNICATION FAILURE ERROR MESSAGES. THIS ERROR RESULTS IN A SYSTEM LOCK UP, NO BOOT, OR SHUT DOWN DEPENDING ON WHEN THE LOSS OF COMMUNICATION OCCURRED. THERE IS NO REPORT OF A PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
176505 6800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS 6800

Patients

Seq Age Sex Outcome Treatment
1