FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ SCHANNEL 3.01
MDR report key: 3823360
·
Received March 20, 2014
Report
- Report Number
- 9615050-2014-02196
- Event Type
- Malfunction
- Date Received
- March 20, 2014
- Date of Event
- January 1, 2014
- Report Date
- February 20, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING IT WAS NOTED THE FAN WAS NOT TURNING. THE PROBABLE CAUSE OF THE CUSTOMER REPORTED S233 WAS BROKEN FAN. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THE DEVICE ALARMED WITH A S233 (OVER TEMPERATURE ERROR) MALFUNCTION ALARM CODE. THERE WERE NO REPORTED ADVERSE EVENTS OR DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 167880 | SYMBIQ SCHANNEL 3.01 | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |