FDA Adverse Event
Malfunction
Summary report: N
MAESTRO MED STEERING DURAGUARD
MDR report key: 3823356
·
Received May 21, 2014
Report
- Report Number
- 0001811755-2014-01848
- Event Type
- Malfunction
- Date Received
- May 21, 2014
- Date of Event
- April 23, 2014
- Report Date
- April 23, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HBB
- PMA / PMN Number
- K041754
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DURING FAILURE ANALYSIS, THE REPORTED EVENT OF A BENT DURAGUARD FOOT WAS CONFIRMED BY THE TECHNICIAN THROUGH VISUAL EVALUATION. A BENT DURAGUARD CAN OCCUR IF EXCESSIVE SIDE LOAD WAS APPLIED TO THE FEATURE. THIS IS NOT A REPAIRABLE DEVICE; THEREFORE, IT WAS NOT RETURNED TO THE USER FACILITY FOLLOWING EVALUATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE FOOT OF THE MAESTRO MEDIUM STEERING DURAGUARD WAS BENT DURING SERVICE INSPECTION AT THE MANUFACTURING FACILITY. NO PATIENT INVOLVEMENT, NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION, AND NO SURGICAL DELAY WERE REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 303896 | MAESTRO MED STEERING DURAGUARD | MOTOR, DRILL, PNEUMATIC | HBB | STRYKER INSTRUMENTS-KALAMAZOO | 07355 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |