FDA Adverse Event Malfunction Summary report: N

ADVANCE BED

MDR report key: 3823342 · Received February 28, 2014

Report

Report Number
1824206-2014-00681
Event Type
Malfunction
Date Received
February 28, 2014
Date of Event
January 31, 2014
Report Date
January 31, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND BROKEN WELDS ON THE RIGHT SIDE RAIL. PER THE HILL-ROM SERVICE MANUAL THE ADVANCE SERES BED REQUIRES EFFECTIVE MAINTENANCE PROGRAM. PREVENTATIVE MAINTENANCE WILL MINIMIZE DOWNTIME DUE TO EXCESSIVE WEAR. TEST THE SIDE RAIL FOR PROPER LATCHING. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2013. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE RIGHT SIDE RAIL TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE TECHNICIAN REPORTED THE RIGHT SIDE RAIL WILL NOT LATCH. THE BED IS LOCATED IN THE BED SHOP AT THE FACILITY. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
125362 ADVANCE BED A/C POWERED ADJUSTABLE BED FNL HILL-ROM, INC. 1105

Patients

Seq Age Sex Outcome Treatment
1