FDA Adverse Event
Malfunction
Summary report: N
ENTERALITE INFINITY ENTERAL FEEDING PUMP
MDR report key: 3823318
·
Received February 28, 2014
Report
- Report Number
- 1722139-2014-00054
- Event Type
- Malfunction
- Date Received
- February 28, 2014
- Date of Event
- January 1, 2013
- Report Date
- January 30, 2014
- Manufacturer
- MOOG MED DEVICES GROUP
- Product Code
- LZH
- PMA / PMN Number
- K981816
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LH
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION BY (B)(4) SERVICE CENTER IN (B)(4) FOUND THAT PUMP EVENT LOG HAD ERROR CODE 13. PUMP PCB BOARD WAS REPLACED. THIS MDR IS BEING SUBMITTED BECAUSE THIS DEVICE IS SIMILAR TO A DEVICE MARKETED IN THE UNITED STATES.
Description of Event or Problem · 1
INFO RECEIVED FROM (B)(4) SERVICE CENTER IN (B)(4) INDICATES THAT PUMP EXPERIENCES ERROR CODE 13.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 124986 | ENTERALITE INFINITY ENTERAL FEEDING PUMP | LZH | MOOG MED DEVICES GROUP | FLOCARE PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |