FDA Adverse Event Malfunction Summary report: N

SYMBIQ DUAL CHANNEL

MDR report key: 3823311 · Received March 20, 2014

Report

Report Number
9615050-2014-02165
Event Type
Malfunction
Date Received
March 20, 2014
Date of Event
February 3, 2014
Report Date
February 4, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT DURING PREVENTATIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE DID NOT PASS THE NOISE THRESHOLD DURING THE TOUCHSCREEN TEST. THIS IS NOT A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, THE DEVICE TOUCHSCREEN DID NOT RESPOND WHEN PRESSED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
167355 SYMBIQ DUAL CHANNEL 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA