FDA Adverse Event
Malfunction
Summary report: N
STENOSCOP
MDR report key: 3823288
·
Received March 20, 2014
Report
- Report Number
- 1720753-2014-02514
- Event Type
- Malfunction
- Date Received
- March 20, 2014
- Date of Event
- March 6, 2014
- Report Date
- March 20, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE CONNECTIONS ON MEMORY CARD AND ACQUISITION CARD WERE EVALUATED AND RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THE LEFT MONITOR INTERMITTENTLY FAILED TO DISPLAY A FLUOROSCOPIC IMAGE. THIS ISSUE WILL CAUSE THE SYSTEM TO BE UNUSABLE DUE TO THE INABILITY TO SEE THE LIVE IMAGE. NO PATIENT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 167732 | STENOSCOP | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | STENOSCOP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |