FDA Adverse Event Malfunction Summary report: N

STENOSCOP

MDR report key: 3823288 · Received March 20, 2014

Report

Report Number
1720753-2014-02514
Event Type
Malfunction
Date Received
March 20, 2014
Date of Event
March 6, 2014
Report Date
March 20, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE CONNECTIONS ON MEMORY CARD AND ACQUISITION CARD WERE EVALUATED AND RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THE LEFT MONITOR INTERMITTENTLY FAILED TO DISPLAY A FLUOROSCOPIC IMAGE. THIS ISSUE WILL CAUSE THE SYSTEM TO BE UNUSABLE DUE TO THE INABILITY TO SEE THE LIVE IMAGE. NO PATIENT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
167732 STENOSCOP FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) STENOSCOP

Patients

Seq Age Sex Outcome Treatment
1