FDA Adverse Event Malfunction Summary report: N

2600

MDR report key: 3823265 · Received February 19, 2014

Report

Report Number
1720753-2014-01672
Event Type
Malfunction
Date Received
February 19, 2014
Date of Event
February 10, 2014
Report Date
February 19, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE CONNECTIONS ON THE CONTROL PANEL PROCESSOR PCB ASSEMBLY WERE EVALUATED AND RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WAS A COMMUNICATION FAILURE ERROR MESSAGE. THIS ERROR IS LIKELY TO RESULT IN A SYSTEM LOCK UP, NO BOOT, OR SHUT DOWN DEPENDING ON WHEN THE LOSS OF COMMUNICATION OCCURRED. THERE IS NO REPORT OF A PATIENT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
105390 2600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS 2600

Patients

Seq Age Sex Outcome Treatment
1