FDA Adverse Event Malfunction Summary report: N

ON-Q PAINBUSTER: SOAKER 5IN: 270ML, 5ML/HR

MDR report key: 3823257 · Received February 19, 2014

Report

Report Number
2026095-2014-00019
Event Type
Malfunction
Date Received
February 19, 2014
Date of Event
December 5, 2013
Report Date
January 22, 2014
Manufacturer
I-FLOW CORP.
Product Code
MEB
PMA / PMN Number
K063530
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

METHOD: AT THIS TIME I-FLOW IS PENDING THE RETURN OF THE DEVICE. ONCE RECEIVED AN INVESTIGATION AND EVALUATION WILL BE PERFORMED ON THE SAMPLE RETURNED. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS CONDUCTED FOR THE LOT NUMBER PROVIDED. RESULTS: RESULTS OF THE EVALUATION ARE IN PROGRESS. PER THE DHR THE LOT MEET MANUFACTURING SPECIFICATIONS AT RELEASE. CONCLUSIONS: A CONCLUSION IS NOT YET AVAILABLE AS THE SAMPLE IS PENDING INVESTIGATION AND EVALUATION ONCE RECEIVED. A FOLLOW-UP REPORT WILL BE FILED WHEN THE INVESTIGATION HAS BEEN COMPLETED. INFORMATION FROM THIS INCIDENT WILL BE INCLUDED IN OUR PRODUCT COMPLAINT AND MDR TREND REPORTING SYSTEM. ADDITIONAL INVESTIGATION MAY ARISE FROM ONGOING ANALYSIS, TREND INFORMATION, OR OTHER ANALYSIS AS APPROPRIATE.

Description of Event or Problem · 1

DRUG/DILUENT: 0.2% ROPIVACAINE. FILL VOLUME: 270ML. FLOW RATE: 5ML/HR. PROCEDURE: POPLITEAL BLOCK. CATHPLACE: UNKNOWN. AN INCIDENT OF FAST FLOW WAS REPORTED. INCIDENT OF FAST FLOW WAS REPORTED. INCIDENT AS REPORTED BY THE DISTRIBUTION: DEFECT CODE- "EMPTY TOO FAST- 24 INSTEAD OF 40, FUNCTION TOO HIGH" INCIDENT AS REPORTED BY THE CUSTOMER: "THE ELASTOMERIC PUMP WAS FILLED WITH 200ML NACI AND ROPIVACAINE 2% (200ML INSTEAD OF 270ML). EVEN IF ONE TAKES ACCOUNT OF THE FACT THAT ONLY 200ML IN 270ML CAUSES MORE RAPID INFUSION, IT IS UNACCEPTABLE FOR THE PUMP TO HAVE EMPTIED IN JUST 24 HOURS RATHER THAN 40 HOURS. APPARENTLY THE CAPILLARY WAS IN PERMANENT CONTACT WITH THE SKIN. PLEASE HAVE THE PUMP ANALYSED AND CHECKED FOR DEFECTS. A NUMBER OF INCIDENTS LIKE THIS HAVE BEEN REPORTED IN RECENT WEEKS." NO ADVERSE EVENTS OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104664 ON-Q PAINBUSTER: SOAKER 5IN: 270ML, 5ML/HR ELASTROMERIC PUMP MEB I-FLOW CORP. PS12504 0200704363

Patients

Seq Age Sex Outcome Treatment
1