FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3823233 · Received March 20, 2014

Report

Report Number
1720753-2014-02539
Event Type
Malfunction
Date Received
March 20, 2014
Date of Event
February 24, 2014
Report Date
March 20, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE BATTERY, FILAMENT DRIVE BOARD, AND THE GENERATOR DRIVE BOARD WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THE SYSTEM HAD A PRECHARGE ERROR. THE PRECHARGE VOLTAGE ERROR WILL LIKELY PREVENT THE SYSTEM FROM BOOTING. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
167344 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1