FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3823233
·
Received March 20, 2014
Report
- Report Number
- 1720753-2014-02539
- Event Type
- Malfunction
- Date Received
- March 20, 2014
- Date of Event
- February 24, 2014
- Report Date
- March 20, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE BATTERY, FILAMENT DRIVE BOARD, AND THE GENERATOR DRIVE BOARD WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THE SYSTEM HAD A PRECHARGE ERROR. THE PRECHARGE VOLTAGE ERROR WILL LIKELY PREVENT THE SYSTEM FROM BOOTING. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 167344 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |