FDA Adverse Event Malfunction Summary report: N

ADVANCE BED

MDR report key: 3823222 · Received February 27, 2014

Report

Report Number
1824206-2014-00629
Event Type
Malfunction
Date Received
February 27, 2014
Date of Event
January 29, 2014
Report Date
January 29, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND BRAKE AND BRAKE STEER CASTERS ARE WORN. THE TECH REPLACED THE BRAKE STEER CASTERS AND BRAKE CASTERS TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE ACCOUNT REPORTED THE BRAKES ARE NOT HOLDING, THE BED STILL ROLLS WHEN THE BRAKE IS SET. THE BED IS LOCATED IN THE STORAGE AREA AT THE FACILITY. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
122631 ADVANCE BED A/C POWERED ADJUSTABLE HOSP BED FNL HILL-ROM, INC. 1165

Patients

Seq Age Sex Outcome Treatment
1