FDA Adverse Event Malfunction Summary report: N

BEMIS 1200CC HIFLOW SUCTION CANISTER

MDR report key: 3823211 · Received February 27, 2014

Report

Report Number
2133713-2014-00006
Event Type
Malfunction
Date Received
February 27, 2014
Date of Event
January 30, 2014
Report Date
February 27, 2014
Manufacturer
BEMIS MFG. CO.
Product Code
GCX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IN RARE INSTANCES, THE FLOAT MECHANISM THAT PREVENTS FLUID FROM ENTERING THE VACUUM SYSTEM WHEN THE CANISTER IS FILLED MAY BECOME STUCK IN THE CLOSED POSITION IN SHIPPING OR STORAGE. THIS WILL PREVENT VACUUM FROM BUILDING IN THE CANISTER. WHILE A PERMANENT CORRECTIVE ACTION IS BEING INVESTIGATED, WE ARE ADVISING CUSTOMERS TO CHECK OUT THE CANISTER PRIOR TO USE, AND ADVISING A METHOD TO FREE THE FLOAT. NOTIFICATION LETTER TO OUR CUSTOMERS, A LEAFLET INCLUDED WITH THE PRODUCT, AND THE RESULTS OF OUR INVESTIGATION ARE INCLUDED WITH THIS REPORT.

Description of Event or Problem · 1

COMPLAINANT STATES THEY ARE HAVING RECURRENT ISSUES WITH THE VACUUM SHUTOFF VALVE STUCK IN THE UPPER OR OFF POSITION WHEN SETTING UP THE CANISTER. WHEN THAT OCCURS, THERE IS NO SUCTION IN THE CANISTER. STAFF MUST REMOVE THE VACUUM SOURCE AND TAP THE CANISTER COVER TO ENABLE THE VALVE TO DROP TO THE LOWER OR ON POSITION. THERE WAS NO PT INVOLVEMENT IN ANY OF THE INCIDENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
122757 BEMIS 1200CC HIFLOW SUCTION CANISTER APPARATUS, SUCTION, VACUUM POWERED GCX BEMIS MFG. CO. 484410 20131206

Patients

Seq Age Sex Outcome Treatment
1