BEMIS 1200CC HIFLOW SUCTION CANISTER
Report
- Report Number
- 2133713-2014-00006
- Event Type
- Malfunction
- Date Received
- February 27, 2014
- Date of Event
- January 30, 2014
- Report Date
- February 27, 2014
- Manufacturer
- BEMIS MFG. CO.
- Product Code
- GCX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
IN RARE INSTANCES, THE FLOAT MECHANISM THAT PREVENTS FLUID FROM ENTERING THE VACUUM SYSTEM WHEN THE CANISTER IS FILLED MAY BECOME STUCK IN THE CLOSED POSITION IN SHIPPING OR STORAGE. THIS WILL PREVENT VACUUM FROM BUILDING IN THE CANISTER. WHILE A PERMANENT CORRECTIVE ACTION IS BEING INVESTIGATED, WE ARE ADVISING CUSTOMERS TO CHECK OUT THE CANISTER PRIOR TO USE, AND ADVISING A METHOD TO FREE THE FLOAT. NOTIFICATION LETTER TO OUR CUSTOMERS, A LEAFLET INCLUDED WITH THE PRODUCT, AND THE RESULTS OF OUR INVESTIGATION ARE INCLUDED WITH THIS REPORT.
COMPLAINANT STATES THEY ARE HAVING RECURRENT ISSUES WITH THE VACUUM SHUTOFF VALVE STUCK IN THE UPPER OR OFF POSITION WHEN SETTING UP THE CANISTER. WHEN THAT OCCURS, THERE IS NO SUCTION IN THE CANISTER. STAFF MUST REMOVE THE VACUUM SOURCE AND TAP THE CANISTER COVER TO ENABLE THE VALVE TO DROP TO THE LOWER OR ON POSITION. THERE WAS NO PT INVOLVEMENT IN ANY OF THE INCIDENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 122757 | BEMIS 1200CC HIFLOW SUCTION CANISTER | APPARATUS, SUCTION, VACUUM POWERED | GCX | BEMIS MFG. CO. | 484410 | 20131206 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |