FDA Adverse Event
Malfunction
Summary report: N
2800
MDR report key: 3823206
·
Received March 20, 2014
Report
- Report Number
- 1720753-2014-02567
- Event Type
- Malfunction
- Date Received
- March 20, 2014
- Date of Event
- March 3, 2014
- Report Date
- March 20, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE AEC BOARD WAS EVALUATED AND REPLACED, AND THE AEC FILES WERE RELOADED. THE KV CONTROL BOARD WAS PROGRAMMED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WAS UNABLE TO PERFORM FLUOROSCOPY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.