FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3823199 · Received March 20, 2014

Report

Report Number
1720753-2014-02550
Event Type
Malfunction
Date Received
March 20, 2014
Date of Event
March 4, 2014
Report Date
March 20, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE DISPLAY ADAPTER PCB AND IMAGE PROCESSOR PCB WERE EVALUATED AND RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM FAILED TO BOOT TO A USABLE STATE AND EXHIBITED A NON-RECOVERABLE LOSS OF FUNCTIONALITY. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
167719 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1