FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3823197 · Received March 20, 2014

Report

Report Number
1720753-2014-02553
Event Type
Malfunction
Date Received
March 20, 2014
Date of Event
March 5, 2014
Report Date
March 20, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE LEFT MONITOR WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE FLUOROSCOPIC IMAGE WAS INTERMITTENTLY LOST FROM THE LEFT "LIVE" MONITOR. THIS ISSUE WILL EFFECTIVELY ELIMINATE THE ABILITY TO VIEW A REAL TIME IMAGE. NO PATIENT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
167722 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1