FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3823141 · Received May 1, 2014

Report

Report Number
3008642652-2014-01201
Event Type
Malfunction
Date Received
May 1, 2014
Date of Event
April 3, 2014
Report Date
April 17, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF CHARGER/MODEM (B)(4) HAS BEEN COMPLETED. UPON EVALUATION, THE CHARGER/MODEM WOULD NOT POWER ON. THE CAUSE OF THE INABILITY TO POWER ON WAS A FLASH MEMORY FAILURE (COMPONENTS U102 AND U105) ON THE C/A BOARD (B)(4). THE FLASH MEMORY HAD AN INTERMITTENT BGA CONNECTION, WHICH WAS DISCOVERED THROUGH THE USE OF A TARGET MEMORY TEST. THE INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURED SOLDER CONNECTION INDUCED BY MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED BUT THE FRACTURED CONNECTION MAY HAVE BEEN ACCELERATED THROUGH ROUGH HANDLING. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDIER CONNECTIONS. A MECHANICAL DESIGN CHANGE TO REDUCE THE PCA STRAIN ((B)(4)) WAS APPROVED BY FDA ON APRIL 16, 2013. IMPLEMENTATION BEGAN ON (B)(4) 2013. RESULTS WILL BE MONITORED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY BOARD IN THE CHARGER/MODEM. THE LAST PATIENT TO USE THIS CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, CHARGER/MODEM (B)(4) WOULD NOT POWER ON. THE LAST PATIENT TO USE THIS CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
262499 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 N/A

Patients

Seq Age Sex Outcome Treatment
1 NA